Analytical solution for contactless electrical impedance measurement Dokunmasiz yolla elektriksel i̇letkenlik ölçümleri için analitik formülasyon

2010-07-15
Top, Can Bariş
Gençer, Nevzat Güneri
It is possible to measure tissue impedance by inducing eddy currents in the tissue, and measuring the magnetic fields caused by these induced eddy currents. In the presented work, an analytical solution is developed for the induced voltage in the receiver coil(s) in the presence of a small volume inhomogeneity in an otherwise homogeneous tissue. Analytical solutions are compared with numerical FDTD (Finite Difference Time Domain) simulations. The differences in magnetic flux were up to 23% when inhomogeneity was located 10 mm deep, whereas that difference was 6% for the inhomogeneity depth of 30 mm. ©2010 IEEE.
2010 15th National Biomedical Engineering Meeting, BIYOMUT2010
Citation Formats
C. B. Top and N. G. Gençer, “Analytical solution for contactless electrical impedance measurement Dokunmasiz yolla elektriksel i̇letkenlik ölçümleri için analitik formülasyon,” presented at the 2010 15th National Biomedical Engineering Meeting, BIYOMUT2010, Antalya, Türkiye, 2010, Accessed: 00, 2023. [Online]. Available: https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=77954442362&origin=inward.