Show/Hide Menu
Hide/Show Apps
Logout
Türkçe
Türkçe
Search
Search
Login
Login
OpenMETU
OpenMETU
About
About
Open Science Policy
Open Science Policy
Open Access Guideline
Open Access Guideline
Postgraduate Thesis Guideline
Postgraduate Thesis Guideline
Communities & Collections
Communities & Collections
Help
Help
Frequently Asked Questions
Frequently Asked Questions
Guides
Guides
Thesis submission
Thesis submission
MS without thesis term project submission
MS without thesis term project submission
Publication submission with DOI
Publication submission with DOI
Publication submission
Publication submission
Supporting Information
Supporting Information
General Information
General Information
Copyright, Embargo and License
Copyright, Embargo and License
Contact us
Contact us
Implant Body Fracture Due to Fatigue in a Maxillary Implant-Retained Overdenture Treatment
Date
2025-01-01
Author
TERZİOĞLU, SEDAT HAKAN
Sarıkaya, Can Hakan
Gürbüz, Rıza
KARAAĞAÇLIOĞLU, LALE
Metadata
Show full item record
This work is licensed under a
Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License
.
Item Usage Stats
31
views
0
downloads
Cite This
The aim of this study was to determine the causes of fatigue and implant failures in an implant-retained overdenture case. One out of four implants portrayed inadequate osseointegration during healing abutment placement, and one was fractured from the middle to the apical third after 1 year of denture use. Various aspects, such as insufficient cooling while implant socket preparation, incorrect prosthodontic planning, and erroneous design of the denture, were evaluated. Macroscopic and microscopic inspections of the fractured implant body were assessed by scanning electron microscopy (SEM) and optical microscopy. Consequently, metallic fatigue along with secondary cracks and beach marks were evident. It was deduced that the failed and fractured implants depicted substandard osseointegration; thus, they were susceptible to the wear out effect of the occlusal forces. Unfavorable osseointegration caused the fulcrum axis to move apically instead of forming on the neck part of the implant, and micromovements around the fulcrum axis mustered the fracture.
Subject Keywords
fatigue
,
fracture
,
implant
,
micromovement
,
overdenture
URI
https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=105000470470&origin=inward
https://hdl.handle.net/11511/114396
Journal
Case Reports in Dentistry
DOI
https://doi.org/10.1155/crid/4064452
Collections
Graduate School of Natural and Applied Sciences, Article
Citation Formats
IEEE
ACM
APA
CHICAGO
MLA
BibTeX
S. H. TERZİOĞLU, C. H. Sarıkaya, R. Gürbüz, and L. KARAAĞAÇLIOĞLU, “Implant Body Fracture Due to Fatigue in a Maxillary Implant-Retained Overdenture Treatment,”
Case Reports in Dentistry