Standard operating procedure for the analysis of major ions in hydrothermal fluids by ion chromatography

2024-01-01
Correggia, Monica
Di Iorio, Luciano
Bastianoni, Alessia Benedicta
Yücel, Mustafa
Cordone, Angelina
Giovannelli, Donato
This standard operating procedure (SOP) describes an ion chromatography (IC) procedure for the major cations and anions in hydrothermal fluids. Hydrothermal fluids are aqueous solutions with a wide range of temperature, salinity, pH and ion species that can be used by microbial metabolism as electron donors and electron acceptors. Due to the high variability of the environmental physical-chemical parameters in these samples, we have developed this protocol taking into account the special features of the matrices analyzed. An Eco IC Metrohm system equipped with a conductivity detector was used. Calibration curves are linear in the 0.1 to 10 mg/L concentration range for cations Ca 2+, Na +, K +, Mg 2+, NH 4+ and anions Cl -, Br -, NO 3-, NO 2-, SO 42-, PO 43-.
Open Research Europe
Citation Formats
M. Correggia, L. Di Iorio, A. B. Bastianoni, M. Yücel, A. Cordone, and D. Giovannelli, “Standard operating procedure for the analysis of major ions in hydrothermal fluids by ion chromatography,” Open Research Europe, vol. 3, pp. 0–0, 2024, Accessed: 00, 2024. [Online]. Available: https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85203555540&origin=inward.