Show/Hide Menu
Hide/Show Apps
Logout
Türkçe
Türkçe
Search
Search
Login
Login
OpenMETU
OpenMETU
About
About
Open Science Policy
Open Science Policy
Open Access Guideline
Open Access Guideline
Postgraduate Thesis Guideline
Postgraduate Thesis Guideline
Communities & Collections
Communities & Collections
Help
Help
Frequently Asked Questions
Frequently Asked Questions
Guides
Guides
Thesis submission
Thesis submission
MS without thesis term project submission
MS without thesis term project submission
Publication submission with DOI
Publication submission with DOI
Publication submission
Publication submission
Supporting Information
Supporting Information
General Information
General Information
Copyright, Embargo and License
Copyright, Embargo and License
Contact us
Contact us
Reduction of false arrhythmia alarms on patient monitoring systems in intensive care units by using fuzzy logic algorithms
Download
index.pdf
Date
2018
Author
Yanar, Erdem
Metadata
Show full item record
Item Usage Stats
192
views
100
downloads
Cite This
Generally in hospitals, monitoring devices in the intensive care units (ICU) have high rates of false arrhythmia alarms independent of their brands and prices. These falsely issued alarms have financial and physiological effects such as redundant usage of hospital resources and hassling patients’ rest, reducing sensitivity of the hospital staff to potential emergency cases, which is named as “false alarm fatigue”. According to Deshmane et al. (2009), 43% of arrhythmia alarms in ICUs are false. Moreover, This rate reaches 90% in some of the arrhythmia types. In our study, we considered that the alarms are triggered by five life threatening conditions, which are asystole (ASY), bradycardia (EBR), tachycardia (ETC), ventricular tachycardia (VTA), ventricular flutter/fibrillation (VFB). These alarms are usually triggered by analysis of ECG and pulsatile waveforms recorded by patient monitoring equipments, which have standard alarm triggering criteria such as instantaneous thresholds on the predictor values. Most of the ICU false alarms are caused by single channel artifacts. In this study, we aim to fuse ECG features with information from other independent signals and get more robust alarm algorithms for ICUs. Pulsatile waveforms, which are highly correlated signals, can be used to corroborate the alarm category and to suppress significant number of false ECG alarms in ICUs. Photoplethysmogram (PPG), arterial blood pressure (ABP) or both PPG, and ABP can be used for this purpose. These waveforms are the least noisy pressure signals available in certain ICUs, and rarely contain ECG-related artifacts. We implemented four different algorithms that use information from ECG, PPG and ABP waveforms. We trained and tested these algorithms on Physionet Challenge 2015 database, which consists of 5 main arrhythmia types and total of 750 recordings. These algorithms have main analysis steps as: pre-processing (bandpass filters to remove baseline artifacts, scaling to normalize the amplitude of waveforms), beat detection, alarm decision (for the generic algorithm). Our results show that if we use only ECG data of the whole dataset, we can obtain 88.3% sensitivity and 77.4% specificity with negligible difference in results between two simultaneous ECG channels. When we use ECG with ABP and PPG combinations, our sensitivity was increased by 8% but specificity decreased by 4%. When we use ECG with PPG combinations, our sensitivity was increased by 6.7% but specificity decreased by 7.9%. These improved methods obtained in this work are around the tolerances accepted by expert physicians, and slightly outperform the results of EBR and VFB cases by the other known algorithms evaluated with the same database.
Subject Keywords
Electrocardiography.
,
Intensive care units.
,
Blood pressure.
,
False alarms.
,
Errors Warnings.
,
Arrhythmia.
URI
http://etd.lib.metu.edu.tr/upload/12621869/index.pdf
https://hdl.handle.net/11511/27102
Collections
Graduate School of Natural and Applied Sciences, Thesis
Suggestions
OpenMETU
Core
Reducing False Arrhythmia Alarms of Patient Monitoring Systems in Intensive Care Units
Yanar, Erdem; Serinağaoğlu Doğrusöz, Yeşim (2017-01-01)
Generally in hospitals, intensive care units (ICUs) have high rates of false arrhythmia alarms independent of their brands and prices. These falsely issued alarms disrupt patients' rest, drain hospital resources, and desensitize the hospital staff to potential emergency situations, which is named as "false alarm fatigue". It has been estimated that 43% of life threatening electrocardiogram (ECG) alarms issued by bedside monitors are false, with some categories of alarm being as high as 90%. In our study, we...
False Ventricular-Fibrillation/Flutter Alarm Reduction of Patient Monitoring Systems in Intensive Care Units
Yanar, Erdem; Serinağaoğlu Doğrusöz, Yeşim (2018-08-16)
Generally in hospitals, false arrhythmia alarm rates are very high in intensive care units (ICUs) patient monitors independent of their brands and prices. These falsely issued alarms disrupt patients' rest, drain hospital resources, and desensitize the hospital staff to potential emergency situations, which is named as false alarm fatigue. It has been estimated that 43% of life threatening electrocardiogram (ECG) alarms issued by bedside monitors are false, with some categories of alarms being as high as 90...
False Arrhythmia Alarms of Patient Monitoring Systems in Intensive Care Units
Yanar, Erdem; Serinağaoğlu Doğrusöz, Yeşim (2017-09-27)
Generally in hospitals, intensive care units (ICUs) have high rates of false arrhythmia alarms independent of their brands and prices. These falsely issued alarms disrupt patients rest, drain hospital resources, and desensitize the hospital staff to potential emergency situations, which is named as false alarm fatigue. It has been estimated that 43% of life threatening electrocardiogram (ECG) alarms issued by bedside monitors are false, with some categories of alarm being as high as 90%. In our study, we co...
Monitoring radiation induced alterations in biological systems, from molecules to tissues, through infrared spectroscopy
Demir, Pinar; Severcan, Feride (2016-01-01)
Humans can be exposed to non-ionizing and ionizing radiation for diagnostic, therapeutic, accidental, and occupational reasons. Consequently, the effect of radiation on biological systems has attracted the attention of researchers for a rather long time. This review is about the mid-infrared Fourier Transform Infrared (FTIR) spectroscopic characterization of non-ionizing and ionizing radiation-induced changes in DNA, lipids, and proteins, as isolated or synthetic macromolecules, and in biological membranes,...
Achieving Clinical Statement Interoperability Using R-MIM and Archetype-Based Semantic Transformations
Kılıç, Özgün Ozan (2009-07-01)
Effective use of electronic healthcare records (EHRs) has the potential to positively influence both the quality and the cost of health care. Consequently, sharing patient's EHRs is becoming a global priority in the healthcare information technology domain. This paper addresses the interoperability of EHR structure and content. It describes how two different EHR standards derived from the same reference information model (RIM) can be mapped to each other by using archetypes, refined message information mode...
Citation Formats
IEEE
ACM
APA
CHICAGO
MLA
BibTeX
E. Yanar, “Reduction of false arrhythmia alarms on patient monitoring systems in intensive care units by using fuzzy logic algorithms,” M.S. - Master of Science, Middle East Technical University, 2018.