Integrated circuit design for flip-chip bonded capacitive micromachined ultrasonic transducers

Download
2013
Maadi, Muhammad
In previous decades, the applications of ultrasound technologies in medical imaging and therapeutic systems have significantly increased. In conventional ultrasound systems, the transducer array is separated from the electronic instrumentation with multicore physical cabling. However, connection cables make the system too bulky and degrade the receive sensitivity in ultrasound 3D imaging applications because of the capacitance of long cables. The interface electronics for phased array ultrasound systems (imaging or therapeutic) use the ultrasound transducer array to steer a focused beam over the volume being fired or imaged with eliminating bulky cables. Among the available types of ultrasound transducers, CMUTs (capacitive micromachined ultrasonic transducers) are particularly interested because they do not suffer from self-heating effects in comparison with their piezoelectric counterparts. Hence, the integration of CMUTs with front-end electronics can be used in high power and continuous wave (CW) applications such as high intensity focused ultrasound (HIFU). The aim of this study is to design highly flexible and programmable transmit beam-former ASIC using a HV 0.35 µm CMOS technology to be flip-chip bonded to a 4X4 CMUT array for ultrasound therapeutic applications. However, proposed IC can be used as a transmitter circuitry in color Doppler 3D imaging applications. In our designed chip, each CMUT element is provided by an 8-bit shift register, an 8-bit comparator, a one-shot circuit with adjustable pulse width, a programmable pulse train generator and a high voltage pulser circuit. The interface electronics can generate two types of outputs with programmable focusing delays to 16 ultrasound transducer elements in different modes: The first mode is for generating single pulses in which the one-shot circuits adjust the width of the pulses from a few nanoseconds to 650 ns with enough resolution for different operating frequencies of ultrasound transducer. The second one has been considered for generating controllable pulse trains. The frequency of the generated pulse trains can be selected using digitally controlled oscillator (DCO) with three 5-bit digitally controlled delay elements (DCDE), frequency down conversion (FDC) circuit and combinational logics. In DCDE circuit depending on the 5-bit digital input vector, 32 different delay settings in the range of 5 to 50 ns can be obtained. For easy prediction of the DCDE delay time for a given digital input vector, we tried to generate a monotonic delay behavior with ascending binary input patterns in different small areas. Since, the stability of designed DCO-DCDE circuit is very important, the circuit was designed stable in -10% VDD and in 25, 50 and 75 °C temperatures variations. The temperature variation values for 25 to 50 °C and 25 to 75 °C are less than 5%. Furthermore, the circuit showed less than 5% variation for -10% supply reduction. The average power consumption of designed 5-bit DCDE with two buffers is 165 and 844 µw when the circuit generates delays of 21 ns (for generating 20 MHz pulse trains when the digital input vector is 11111) and 1.54 ps (for generating 650 GHz pulse trains when the digital input vector is 00000), respectively. To have a better resolution for the frequency of the pulse trains, we added frequency down conversion (FDC) circuit to DCO-DCDE. The output of the FDC circuit can be selected using an 8X1 multiplexer to generate 256 different pulse train frequencies in the range of 1 to 10 MHz with good resolution. Single pulse and pulse train modes are separated from each other by a 1X2 demultiplexer. At final stage the amplitudes of single pulses or pulse trains are increased up to 45 V using high voltage (HV) pulsers with THKOX module. The outputs of these pulsers are connected to flip-chip bonding pads and then the top electrodes of the CMUTs. When the magnitude of the pulses is 45 V, the rise and fall time of the output signals are 5 and 8 ns for 2.5 pF capacitive loads, respectively. The slew rate and figure of merit of the circuit at VHV = 45 V is 1.89 V/µs and 0.51 ns/(µm.V), respectively. These kinds of circuits have dramatic power consumptions. Our designed HV pulser circuit consumes 51.66 mw for 45 V output signals. Using an external control system like FPGA, the 8-bit global counter can be incremented from 1 to 256 so that each pulser circuit fires correspondent CMUT element when its stored shift register value is identical with the value of global counter. In the frame of this research, a different kind of ultrasound therapeutic interface electronics for 16X16 CMUT array has been designed in HV 0.35 µm CMOS technology. The proposed ASIC includes the same driver (LV) circuitry but 90 V high voltage pulsers with HVTHKOX module.

Suggestions

Design of intelligent nanoparticles for use in controlled release
Bayyurt, Banu; Hasırcı, Vasıf Nejat; Department of Biotechnology (2009)
The aim of this project was to design an intelligent controlled release system based on thermoresponsive nanoparticles for cancer therapy and to evaluate the efficiencies of these systems with in vitro cell culture. Poly(Nisopropylacrylamide), an important thermoresponsive polymer, was selected for this study to prepare the responsive nanoparticles. This polymer has an lower critical solution temperature (LCST) of 32 oC, below which it is hydrophilic and above this temperature, it shows hydrophobic behavior...
Capacitive micromachined ultrasonic transducer arrays for integrated diagnostic/therapeutic catheters
Wong, Serena H.; Wygant, Ira O.; Yeh, David T.; Zhuang, Xuefeng; Bayram, Barış; Kupnik, Mario; Oralkan, Omer; Ergun, A. Sanli; Yaralioglu, Goksen G.; Khuri-Yakub, Butrus T. (2006-01-01)
In recent years, medical procedures have become increasingly non-invasive. These include endoscopic procedures and intracardiac interventions (e.g., pulmonary vein isolation for treatment of atrial fibrillation and plaque ablation for treatment of arteriosclerosis). However, current tools suffer from poor visualization and difficult coordination of multiple therapeutic and imaging devices. Dual-mode (imaging and therapeutic) ultrasound arrays provide a solution to these challenges. A dual-mode transducer ca...
Integrated circuit design for annular capacitive micromachined ultrasonic transducer
Kılıç, Yusuf Kutluhan; Bayram, Barış; Department of Electrical and Electronics Engineering (2015)
The conventional ultrasound systems consist of many separated parts, and the bulky cables for connections result in a problem to be resolved. The systems with integrated interface electronics preferable since the removal of bulky cables resolves the signal degradation problems. In this thesis, a flexible and high drive capability digitally controlled beamformer is designed. The beamformer is designed to be integrated with annular capacitive micromachined ultrasonic transducer (CMUT) arrays. Although the bea...
Performance evaluation of similarity measures for dense multimodal stereovision
Yaman, Mustafa; Kalkan, Sinan (SPIE-Intl Soc Optical Eng, 2016-05-01)
Multimodal imaging systems have recently been drawing attention in fields such as medical imaging, remote sensing, and video surveillance systems. In such systems, estimating depth has become possible due to the promising progress of multimodal matching techniques. We perform a systematic performance evaluation of similarity measures frequently used in the literature for dense multimodal stereovision. The evaluated measures include mutual information (MI), sum of squared distances, normalized cross-correlat...
Design and verification of diamond based capacitive micromachined ultrasonic transducer
Çetin, Ahmet Murat; Bayram, Barış; Department of Electrical and Electronics Engineering (2011)
Potential applications such as high intensity focused ultrasound in medical therapeutics require larger output pressures. To offer unprecedented acoustic output pressure in transmit without the limitations, Capacitive Micromachined Ultrasonic Transducer operation modes of collapse and collapse-snapback are introduced in literature. Both operation modes require the membrane to contact the substrate surface, which poses a problem on the durability of the membrane in terms of structural integrity and tribologi...
Citation Formats
M. Maadi, “Integrated circuit design for flip-chip bonded capacitive micromachined ultrasonic transducers,” M.S. - Master of Science, Middle East Technical University, 2013.