Design and fabrication of a high performance resonant MEMS temperature sensor

2016-04-01
This paper presents a high performance MEMS temperature sensor comprised of a double-ended-tuning-fork (DETF) resonator and strain-amplifying beam structure. The temperature detection is based on the 'thermal strain induced frequency variations' of the DETF resonator. The major source of thermal strain leading to the frequency shifts is the difference in thermal expansion coefficients of the substrate and the device layers of the fabricated structures. By selecting the substrate as glass and the device layers as single crystal silicon, i.e. materials with different thermal expansion coefficients, the tines of the resonators are exposed to axial load with the changing temperature, which causes a change in the resonance frequency of the resonators. This resonance frequency shift can be related with the changing temperature by taking the thermal strain relations into consideration, which enables utilization of the resonator as a highly sensitive temperature sensor. The resonators used in this study have been fabricated by utilizing the advanced MEMS process that incorporates the simple silicon-on-glass process with the wafer level vacuum packaging Torunbalci et al (2015 J. Microelectromech. Syst. 24 556-64). The fabricated resonators have been tested in a temperature-controlled oven between -20 degrees C and 60 degrees C, and the results of two distinct designs are compared to be able to observe the effectiveness of the strain amplifying beam. Measurement results show that the design with the strain amplifying beam increases the temperature coefficient of frequency of the resonators by 33 times when compared to the one-end free DETF resonators. Minimum detectable temperature variations observed by the resonators used in this study is 0.0011 degrees C. This kind of very high resolution temperature sensing can be achieved by integrating this MEMS temperature sensor with any type of physical MEMS sensor where its fabrication process includes different materials for the substrate and sensor structures.
JOURNAL OF MICROMECHANICS AND MICROENGINEERING

Suggestions

An on-die ultra-low voltage DC-DC step-up converter with voltage doubling LC-tank
Jayaweera, H. M. P. C.; Pathirana, W. P. M. R.; Muhtaroglu, Ali (IOP Publishing; 2016-12-01)
In this paper we report the design, characterization and verification of a novel on-die ultra-low voltage DC-DC converter circuit for energy harvester applications in 0.18 mu m complementary metal oxide semiconductor (CMOS) technology. The circuit self-starts, does not use off-chip components, and is thus suitable for use in highly integrated low cost systems. The first version of the design has a five-stage charge-pump stimulated by an oscillator with two center-tap inductors. It is validated on a test chi...
Modeling and fabrication of electrostatically actuated diaphragms for on-chip valving of MEMS-compatible microfluidic systems
Atik, Ali Can; Ozkan, Metin Dundar; Ozgur, Ebru; Külah, Haluk; Yıldırım, Ender (IOP Publishing, 2020-11-01)
This paper presents an analytical model to estimate the actuation potential of an electrostatic parylene-C diaphragm, processed on a glass wafer using standard microelectromechanical systems (MEMS) process technology, and integrable to polydimethylsiloxane (PDMS) based lab-on-a-chip systems to construct a normally-closed microvalve for flow manipulation. The accurate estimation of the pull-in voltage of the diaphragm is critical to preserve the feasibility of integration. Thus, we introduced an analytical m...
The effects of tine coupling and geometrical imperfections on the response of DETF resonators
Azgın, Kıvanç (IOP Publishing, 2013-12-01)
This paper presents a two-degree-of-freedom analytical model for the electromechanical response of double ended tuning fork (DETF) force sensors. The model describes the mechanical interaction between the tines and allows investigation of the effect of a number of asymmetries, in tine stiffness, mass, electromechanical parameters and load sharing between the tines. These asymmetries are introduced during fabrication (e. g., as a result of undercut) and are impossible to completely eliminate in a practical d...
Analysis and characterization of an electrostatically actuated in-plane parylene microvalve
Yıldırım, Ender; Külah, Haluk (IOP Publishing, 2011-10-01)
This paper presents analysis and implementation of a simple electrostatic microvalve designed for use in parylene-based lab-on-a-chip devices. The microvalve utilizes an in-plane collapsing diaphragm. To investigate the pull-in behavior of the diaphragm and flow characteristics, a thorough analysis is carried out using the finite element method. Microvalves with different diaphragm radii are fabricated using surface micromachining techniques. Pull-in tests are carried out under the no-flow condition with ai...
Variable Temperature-Scanning Hall Probe Microscopy With GaN/AlGaN Two-Dimensional Electron Gas (2DEG) Micro Hall Sensors in 4.2-425 K Range Using Novel Quartz Tuning Fork AFM Feedback
Akram, R.; Dede, A.; Oral, Ahmet (Institute of Electrical and Electronics Engineers (IEEE), 2008-11-01)
In this paper, we present the fabrication and variable temperature (VT) operation of Hall sensors, based on GaN/AlGaN heterostructure with a two-dimensional electron gas (2DEG) as an active layer, integrated with quartz tuning fork (QTF) in atomic force-guided (AFM) scanning Hall probe microscopy (SHPM). Physical strength and a wide bandgap of GaN/AlGaN heterostructure makes it a better choice to be used for SHPM at elevated temperatures, compared to other compound semiconductors (AlGaAs/GaAs and InSb), whi...
Citation Formats
T. Kose, K. Azgın, and T. Akın, “Design and fabrication of a high performance resonant MEMS temperature sensor,” JOURNAL OF MICROMECHANICS AND MICROENGINEERING, pp. 0–0, 2016, Accessed: 00, 2020. [Online]. Available: https://hdl.handle.net/11511/42337.